It's a ceiling-mounted LED shadowless operating lamp designed for use in operating theaters and procedure rooms, including stomatology, cosmetology, ophthalmology and general medical operations. The "ENDO" designation indicates suitability for endoscopic procedures where focused, high-quality illumination is required.
The lamp provides illuminance of ≥120,000 lux (per lamp head) and an illuminance depth of 1045 mm, suitable for deep surgical fields and high‑precision procedures.
Color temperature is adjustable in the range from about 3800 K to 5000 K (specifically 3800±200K, 4000±200K, 4500±200K and 5000±200K). Beam synthesis control technology allows users to change color temperature to optimize tissue contrast.
The lamp offers a very high color rendering ability with a CRI around Ra96, and beam synthesis control ensures strong color fidelity (CR I maintained at or above typical surgical requirements, e.g., ≥93 in adjustable modes).
The illuminant service life is ≥50,000 hours under normal operating conditions.
Each lamp head has a diameter of 700 mm. The adjustable light field size ranges from 220 mm to 310 mm in diameter.
This is a ceiling-mounted unit. The lowest recommended installation height is 2900 mm from the floor. Professional installation by qualified technicians is recommended to meet local codes and theater layout.
The lamp accepts AC 110–240 V at 50/60 Hz. For exact current draw or power consumption details for a specific configuration, consult the supplier or technical datasheet.
Each lens is a 40 mm super-large lens arranged like an insect compound eye with inter-lens spacing under 1 mm to reduce shadows. In deep lighting mode, the system automatically disables ineffective LEDs and selectively boosts effective LEDs to repair localized shadows, providing up to ~20% brighter deep shadowless illumination compared with traditional shadowless lamps.
Total irradiance is specified as 127 W/m². As an LED system, it produces less radiant heat at the surgical site than equivalent halogen lamps, but some heat is still present; theatre ventilation and standard surgical precautions should be maintained.
The light field diameter is adjustable between 220 mm and 310 mm. Color temperature and brightness can also be adjusted via the lamp's control interface to suit the procedure and surgeon preference.
Yes. The product is specified with an "ENDO" designation and is suitable for endoscopic applications thanks to its strong, focused illumination, adjustable light field and high color rendering. For integration with specific endoscopic camera systems, check compatibility with the equipment supplier.
Clean external surfaces with approved hospital disinfectants and soft cloths to avoid scratching lenses. Inspect mounting hardware and electrical connections periodically. LED modules have long service life (≥50,000 h), but follow the manufacturer's maintenance schedule for lamp checks and any component replacement.
The standard specification does not list an integrated battery backup. Emergency or UPS options may be available as accessories or as part of a custom configuration—consult your supplier for backup power solutions and optional features.
Contact the manufacturer or authorized distributor for installation services, warranty terms, local regulatory compliance information and technical support. They can provide detailed installation instructions, mounting requirements and service arrangements for your facility.