·12" color TFT display, up to 12 waveforms · One-level menu, easy to use ·Parameters including ECG/RESP, SpO2 , NIBP, TEMP, Pulse Rate and IBP(optional) ·Maximum 720 hour graphic and tabular trends of all parameters ·Built-in recorder and battery ·Suitable for adult, pediatric and neonatal patient ·Arrhythmia analysis, ST calculation, Pace analysis ·Anti-ESU, anti- defibrillator ·Standard network interface, can connect to Central monitor system ·Optinal: recorder, dual IBP, CO2, IoC
Specification
item
value
Brand Name
Jumper
Model Number
JPD-800B
Specifications
ECG Lead mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V) Lead selection I, II, III, avR, avL, avF, V, Waveform 2 ch Lead mode 3 Leads ( R, L, F or RA, LA, LL) Lead selection I, II, III, Waveform 1 ch Gain: X2.5mm/mV, X5.0mm/mV, X10mm/mV, X20mm/mV, auto Resolution 1 bpm Sensitivity > 200 (uV P-P) Differential Input Impedance > 5 MΩ
CMRR Monitor > 105 dB Operation > 105 dB Diagnosis > 85 dB Electrode offset potential 300mV Leakage Current < 10 uA Baseline Recovery < 3 S After Defi. ECG Signal Range: 8 m V ( Vp-p ) Bandwidth Surgery 1 ~ 15 Hz Monitor 0.5 ~ 35 Hz Diagnostic 0.05 ~ 100 Hz 5%±Calibration Signal 1 (mV p-p), Accuracy : ST Segment Monitoring Range Measure and Alarm -2.0 ~ +2.0 mV ARR DetectingType ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC Alarm Available Review Available
RESPARATION Method Impedance between R-F(RA-LL) Differential Input Impedance >2.5 MΩ Measuring Impedance Range: 0.3~5.0Ω Base line Impedance Range: 0 ~ 2.5 KΩ Bandwidth 0.3 ~ 2.5 Hz Resp. Rate Measuring and Alarm Range Adult 0 ~ 120 rpm Neo/Ped 0 ~ 150 rpm Resolution 1 rpm Accuracy 2 rpm Apean Alarm 10 ~ 40 S
NIBP Method Oscillometric Mode Manual, Auto, STAT Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min) Measuring Period in STATMode 5 Min Pulse Rate Range 40 ~ 240 bpm Alarm Type SYS, DIA, MEAN Measuring and alarm range Adult Mode SYS 40 ~ 270 mmHg DIA 10 ~ 215 mmHg MEAN 20 ~ 235 mmHg Pediatric Mode SYS 40 ~ 200 mmHg DIA 10 ~ 150 mmHg MEAN 20 ~ 165 mmHg Neonatal Mode SYS 40 ~ 135 mmHg DIA 10 ~ 100 mmHg MEAN 20 ~ 110 mmHg Resolution Pressure 1mmHg
Accuracy&Pressure Maximum Mean error 5mmHg Maximum Standard deviation 8mmHg Overpressure Protection Adult Mode 297 3 mmHg Pediatric Mode 240 3 mmHg Neonatal Mode 147 3 mmHg
SpO2 Measuring Range 0 ~ 100 % Alarm Range 0 ~ 100 % Resolution 1 % Accuracy 70% ~ 100% 2 % 0% ~ 69% unspecified Actualization interval about 1 Sec. Alarm Delay 10 Sec. Pulse Rate Measuring and Alarm Range 0~254bpm Resolution 1bpm Accuracy 2bpm
TEMPERATURE Channel 1 Measuring and Alarm Range 0 ~ 50 C Resolution 0.1 C Accuracy 0.1 C Actualization interval about 1 Sec. Average Time Constant < 10 Sec.
IBP(Optional) Label ART, PA, CVP, RAP, LAP, ICP, P1, P2 Measuring and alarm range ART 0 ~ 300 mmHg PA -6 ~ 120 mmHg CVP/RAP/LAP/ICP -10 ~ 40 mmHg P1/P2 -10 ~ 300 mmHg Press Sensor Sensitivity 5 uV/V/mmHg Impedance 300-3000Ω Resolution 1 mmHg Accuracy 2% or 1mmHg, which great Actualization interval about 1 Sec.
Afrimart Online stands as Africa's foremost e-commerce hub, featuring an expansive array of product categories spanning electronics, machinery, toys, furniture, and more. Committed to excellence, our platform delivers swift and streamlined logistics and customs clearance solutions, offering a comprehensive one-stop service for cross-border imports.
Leveraging over three decades of international trade expertise, we ensure a seamless and hassle-free importing experience, providing top-tier products at competitive prices to meet the diverse needs of African consumers.