The AWM43600 Micro-Bridge Air Quality Flow Sensor is primarily designed for leak detection devices in medical applications, providing precise air flow measurement.
The output voltage of the AWM43600 sensor is 5.0 V ±0.15 Vdc at a flow rate of 6 SLPM.
The maximum power consumption of the AWM43600 sensor is 75 mW at an operating temperature of 85 °C.
The AWM43600 sensor can handle a full scale flow range of up to +6 SLPM.
The typical response time of the AWM43600 sensor is 1 ms, with a maximum response time of 3 ms.
The AWM43600 sensor has a repeatability and hysteresis within ±1.00% reading, ensuring reliable performance.
At full scale, the AWM43600 sensor operates under a pressure drop of 8.00 inches of H₂O (typical).
The AWM43600 sensor can tolerate a maximum common mode pressure of 25 psi.
Yes, the AWM43600 sensor is energy efficient, consuming only 75 mW (max) at 85 °C.
Yes, the AWM43600 sensor can sink 10 mA and source 20 mA, supporting integration with various medical and industrial systems.
The AWM43600 sensor is manufactured by Honeywell and its model number is FLOW SENSOR AWM3300V.
The typical applications for the AWM43600 sensor include medical devices for leak detection and industrial applications for environmental monitoring and air quality systems.
The null voltage for the AWM43600 sensor is specified as 1.0 ±0.05 Vdc.
Yes, the AWM43600 sensor has a typical ratiometricity error of ±0.8% reading.
The AWM43600 sensor is packaged using professional, environmentally friendly, and efficient packaging services to ensure the safety of the goods.
Discover our latest orders